Apera Instruments AI531

Apera Instruments PH820 Benchtop pH Meter Kit

Instruction Manual

Продукт бүттүview

The Apera Instruments PH820 Benchtop pH Meter Kit is designed for reliable and accurate pH measurements in laboratory settings. It features advanced functionalities including intelligent calibration, data management, and a durable, water-resistant structure. This manual provides detailed instructions for proper use and care of your PH820 meter.

Apera Instruments PH820 Benchtop pH Meter Kit with electrode and stand

Figure 1: Apera Instruments PH820 Benchtop pH Meter Kit, showing the main unit, electrode, temperature probe, and electrode arm.

Key features of the PH820 include:

  • Жогорку тактык: Achieves ±0.002 pH±1 digit precision.
  • Premium Electrode: Equipped with the Swiss LabSen 211 Glass-body pH Electrode for stable and accurate measurements.
  • Intelligent Calibration: Supports 1 to 5 points auto-calibration with calibration guide and reminder functions.
  • Өзүн-өзү диагностикалоо: Assists in identifying and resolving common calibration issues and monitors electrode health.
  • Маалыматтарды башкаруу: GLP-compliant data storage for up to 500 groups with USB data output and PC software support.
  • Туруктуу дизайн: IP54 water-resistant construction.

Жайгашуу

Пакеттен чыгаруу жана компоненттерди текшерүү

Carefully unpack all components from the kit. Verify that the following items are present:

  • PH820 pH Meter Unit
  • LabSen 211 рН электрод
  • Температура проба
  • Flexible Electrode Arm and Holder
  • USB кабели
  • PC-Link Software Flash Drive
  • Power Adaptor (Input: 120V AC 60Hz; Output: 9V 300mA)
  • Calibration Solutions (4.00 pH, 7.00 pH, 10.01 pH)

Туташтыруучу компоненттер

Connect the electrode, temperature probe, and power adapter to the corresponding ports on the back of the PH820 meter unit. Ensure all connections are secure.

Back panel of Apera Instruments PH820 pH meter showing connection ports

2-сүрөт: Арткы view of the PH820 meter, highlighting the REF, pH/mV, TEMP, RS232, and DC 9V input ports.

  1. Электрод туташуу: Plug the LabSen 211 pH Electrode into the port labeled "pH/mV".
  2. Температуралык зондук туташуу: Insert the temperature probe into the port labeled "TEMP".
  3. Кубат туташуу: Connect the power adapter to the "DC 9V" port and then to a power outlet.
  4. Electrode Arm Assembly: Attach the flexible electrode arm and holder to the side of the meter unit for convenient electrode placement during measurements.

Иштөө нускамалары

Basic pH Measurement

  1. Power on the meter by pressing the power button.
  2. Электродду дистилденген суу менен чайкап, акырын кургатыңыз.
  3. Immerse the electrode and temperature probe into the sample solution. Ensure the electrode tip is fully submerged.
  4. Wait for the reading on the large LCD to stabilize. The meter displays pH and temperature simultaneously.
  5. Record the stable reading.

рН калибрлөө

The PH820 supports 1 to 5 points automatic calibration. Regular calibration ensures measurement accuracy.

  1. Prepare the standard calibration solutions (4.00 pH, 7.00 pH, 10.01 pH are included).
  2. басыңыз КАЛ калибрлөө режимине өтүү үчүн баскыч.
  3. Rinse the electrode and immerse it into the first calibration solution (e.g., 7.00 pH).
  4. The meter will automatically recognize the solution and begin calibration. Wait for the stability icon to appear.
  5. Once stable, the meter will prompt for the next calibration point. Repeat the rinsing and immersion process for subsequent solutions.
  6. The self-diagnosis function will provide feedback on the calibration process and electrode condition, including slope data.
  7. Exit calibration mode once all desired points are completed.

Маалыматтарды башкаруу

The PH820 allows for GLP data management, storing up to 500 groups of data. Data can be output via USB to a PC using the provided PC-Link software (Windows-based system only). Auto-timing data logging can also be configured.

ORP Measurement (Optional)

The PH820 can measure Oxidation-Reduction Potential (ORP) when an ORP electrode (e.g., 301Pt-C, sold separately) is connected to the pH/mV port. Refer to the ORP electrode's specific instructions for calibration and measurement procedures.

Техникалык тейлөө

Электродго кам көрүү

Proper care of the LabSen 211 pH Electrode is crucial for maintaining accuracy and extending its lifespan.

Close-up of the LabSen 211 pH electrode

Сүрөт 3: Жакын пландан view of the LabSen 211 pH Electrode, showing its glass body and sensor tip.

  • Чайкоо: Ар бир өлчөөнүн алдында жана андан кийин электродду дистилденген же деионизацияланган суу менен чайкап туруңуз.
  • Сактагыч: Store the electrode in its protective cap filled with a suitable electrode storage solution (or 3M KCl solution). Never store the electrode dry or in distilled water.
  • Тазалоо: If the electrode becomes contaminated or readings are sluggish, clean it according to the electrode's specific cleaning instructions.
  • Bubble-free Tip: The LabSen 211 features an impact-resistant and bubble-free sensor tip. Ensure no air bubbles are trapped around the sensor during measurement.

Meter Care

The PH820 meter unit is IP54 water-resistant, meaning it is protected from dust and splashing water. However, it should not be submerged. Keep the meter clean and dry. Use a soft, damp cloth to wipe the surface. Avoid using abrasive cleaners or solvents.

Проблемаларды чечүү

The PH820 features a self-diagnosis function that can assist in identifying and fixing common problems, particularly during calibration. Refer to the meter's display for diagnostic messages.

КөйгөйМүмкүн себепЧечим
Туруксуз көрсөткүчтөрDirty electrode; air bubbles; insufficient sample volume; expired calibration.Clean electrode; ensure no bubbles; use adequate sample; recalibrate.
Калибрлөө катасыIncorrect buffer solution; contaminated electrode; electrode aging.Use fresh, correct buffers; clean electrode; check electrode slope data (self-diagnosis) for replacement indication.
Дисплей/Кубат жокPower adapter issue; loose connection.Check power adapter and outlet; ensure all cables are securely connected.
Туура эмес окууларElectrode not calibrated; temperature not compensated; damaged electrode.Perform calibration; ensure temperature probe is functioning; inspect electrode for damage.

For persistent issues not resolved by the above, please contact Apera Instruments customer support.

Техникалык шарттар

ПараметрНарк
Модель номериAI531
рН тактыгы±0.002 pH±1 digit
Калибрлөө пункттары1 to 5 points (Auto.)
Маалыматтарды сактоо500 groups (GLP compliant)
Маалымат чыгарууUSB (PC software support)
Суу каршылыкIP54
Power Input120 В AC 60 Гц
Power Output9V 300mA
Пакеттин өлчөмдөрү10.67 x 10.2 x 6.65 дюйм
Салмагы6.69 фунт
СертификаттарCE, RoHS, ISO 9001:2015

Кепилдик жана колдоо

Apera Instruments products are manufactured under an ISO 9001:2015 certified quality system, ensuring high standards of production. For specific warranty information and technical support, please refer to the warranty card included with your product or visit the official Apera Instruments website. You may also contact their customer service directly for assistance with product operation, troubleshooting, or parts replacement.

© 2024 Apera Instruments. All rights reserved.