1. Киришүү
This manual provides essential information for the safe and effective use of your Senyang Alternating Pressure Anti-Bedsore Air Medical Mattress, Model T05+P06C. This system is designed to aid in the prevention and treatment of pressure ulcers (bedsores) by continuously changing pressure points, promoting blood circulation, and enhancing patient comfort. Please read this manual thoroughly before setup and operation to ensure proper usage and to maximize the benefits of the product.
2. Пакеттин мазмуну
Carefully unpack all components and verify that all items are present and undamaged. If any items are missing or damaged, please contact customer support immediately.
- Air Mattress (80"L x 34"W x 4.5"H, 18 cells)
- Air Pump (Control Unit)
- Ремонт комплекти

3. Техникалык шарттар
The following table outlines the key specifications for the Senyang Alternating Pressure Anti-Bedsore Air Medical Mattress system:
| Өзгөчөлүк | Спецификация |
|---|---|
| Бренд аты | Сеньянг |
| Модель номери | T05+P06C |
| Төшөктүн өлчөмдөрү | 80"L x 34"W x 4.5"H (18 cells) |
| Аба клеткасынын материалы | Medical grade Nylon TPU |
| Мукаба материалы | Stretched PU, waterproof |
| Салмак сыйымдуулугу | 150 кг |
| Насостун көлөмү | 27.5L x 12W x 10H см |
| Pump Cycle Time | 6 / 9 / 12 / 25 minutes (adjustable) |
| Аба чыгаруу | More than 6 Liter/min |
| басым диапазону | 14mmHg - 62mmHg |
| Өзгөчөлүктөрү | Foldable, Care Mattress |
| Колдонмо | Оорукана |
| Кепилдик | Mattress: 1 year, Pump: 2 years |
4. Орнотуу нускамалары
Follow these steps to properly set up your alternating pressure mattress system:
- Төшөктү даярдоо: Place the mattress on a standard hospital bed frame. Ensure the bed surface is clean and free of sharp objects.
- Матрасты жайгаштыруу: Lay the air mattress flat on the bed frame. Ensure the foot end of the mattress aligns with the foot end of the bed. The air hoses should be positioned towards the foot end.
- Secure the Mattress: If your mattress has straps or ties, secure it to the bed frame to prevent shifting.
- Насосту жайгаштырыңыз: Hang the air pump (control unit) securely at the foot end of the bed, typically on the bed rail, using the integrated hooks. Ensure it is stable and not prone to falling.
- Аба шлангдарын туташтыруу: Connect the air hoses from the mattress to the corresponding air outlets on the pump. Ensure connections are firm and secure to prevent air leakage. The image below shows the mattress and pump connected.
- Туташуу кубаты: Насостун шнурун жерге туташтырылган электр розеткасына сайыңыз.
- Initial Inflation: Turn on the pump using the power switch. The mattress will begin to inflate. Allow approximately 20-30 minutes for the mattress to fully inflate and stabilize before placing a patient on it.


5. Иштөө нускамалары
Once the mattress is inflated, you can adjust the settings on the control unit to suit the patient's needs.
5.1 Башкаруу панелинин үстүнөview
The pump's control panel features various buttons and indicators for managing the mattress's functions.

5.2 Негизги функциялар
- Кубат которгучу: Located on the side or back of the pump, turns the unit ON/OFF.
- Басымды жөндөө: Use the pressure control dial or buttons to set the desired pressure range (14mmHg - 62mmHg). Adjust according to patient comfort and clinical guidelines.
- Цикл убактысы: Select the alternating cycle time (6, 9, 12, or 25 minutes) using the dedicated button or dial. This determines how frequently the pressure in the air cells alternates.
- Панелдин кулпусу: Activates or deactivates the control panel lock to prevent accidental changes to settings.
- Static/Initiating Mode: This mode typically provides constant low pressure for patient transfers or when alternating pressure is not desired. 'Initiating' may refer to the initial inflation phase.
- Auto Firm: Provides maximum inflation for a firm surface, useful for patient repositioning or care procedures.
- Ойготкучту баштапкы абалга келтирүү: Silences any active alarms (e.g., low pressure alarm). Address the cause of the alarm before resetting.
5.3 Жалпы операция
- After initial inflation, adjust the pressure setting to a comfortable level for the patient, ensuring adequate support without bottoming out.
- Select an appropriate cycle time. Shorter cycles provide more frequent pressure changes.
- Monitor the patient regularly for comfort and skin integrity. Adjust settings as needed.
- Куч болгон учурда уtage, the mattress will slowly deflate. Ensure alternative care plans are in place.
6. Техникалык тейлөө
Regular maintenance ensures the longevity and optimal performance of your mattress system.
6.1 Тазалоо
- Матрас капкагы: The stretched PU, waterproof cover can be wiped down with a mild disinfectant solution. Ensure the cover is completely dry before use. Do not immerse the mattress or pump in water.
- Аба насосу: Жарнама менен насостун сыртын сүртүңүзamp кездеме. Абразивдүү тазалагычтарды же эриткичтерди колдонбоңуз.
6.2 Leak Checks and Inspection
Periodically inspect the mattress and hoses for any signs of wear, punctures, or leaks.
- Шланг текшерүү: Check air hoses for leaks when they are ready for use.
- Mattress Inflation Check: The manufacturer performs a full check on mattresses with a 24-hour air inflation inspection to ensure quality. Users should also periodically check for consistent inflation.

6.3 Оңдоо комплектин колдонуу
Use the provided repair kit for minor punctures in the air mattress. Follow the instructions included with the repair kit for proper application.
7. Кыйынчылыктарды
If you encounter issues with your mattress system, refer to the following common problems and solutions:
| Көйгөй | Мүмкүн себеп | Чечим |
|---|---|---|
| Насос иштебей жатат | Электр энергиясы жок | Электр шнурунун туташуусун жана розетканы текшериңиз. Кубат которгучу күйүк экенин текшериңиз. |
| Mattress not inflating or losing pressure | Loose hose connection, puncture in mattress/hoses, pump malfunction | Ensure all hose connections are secure. Inspect mattress and hoses for visible damage; use repair kit for small punctures. If problem persists, contact support. |
| Насос ызы -чуу | Normal operation, pump not placed on a stable surface | Some noise is normal. Ensure pump is securely hung or placed on a flat, stable surface to minimize vibration. |
| Ойготкуч угулат | Low pressure, power failure | Check for leaks or power issues. Press 'Alarm Reset' after resolving the issue. |
8. Кепилдик жана колдоо
Senyang is committed to providing high-quality products and excellent customer service.
8.1 Продукциянын кепилдиги
- Матрас: Сатып алган күндөн тартып 1 жылдык кепилдик.
- Насос: Сатып алган күндөн тартып 2 жылдык кепилдик.
This warranty covers manufacturing defects under normal use. It does not cover damage due to misuse, accidents, or unauthorized modifications.
8.2 Кардарларды колдоо
For any questions regarding the use, maintenance, or troubleshooting of your Senyang mattress system, please contact our customer service team.
- We provide online guidance for product usage inquiries.
- Complaints are handled within 24 hours.
- Within a reasonable range, free parts can be delivered by courier.

8.3 Quality Assurance and Certifications
Our products adhere to strict quality control standards and have obtained various system and quality certificates, including ISO13485, TUV, CE, and 510K, ensuring reliability and safety.