Victor 01S Temperature Source Signal Generator User Manual

Модели: 01S

1. Киришүү

The Victor 01S Temperature Source Signal Generator is a battery-powered, handheld portable instrument designed for precise output of electrical and physical parameters. It is capable of analog output for DC voltage, resistance, thermocouple (TC), and thermal resistance (RTD) signals, making it an essential tool for calibration and testing in various industrial and laboratory settings.

This manual provides detailed instructions for the safe and efficient use of your Victor 01S signal generator. Please read it thoroughly before operation and keep it for future reference.

2. Коопсуздук маалыматы

To ensure safe operation and service of the instrument, follow these safety precautions:

  • Always inspect the instrument and test leads for damage before use. Do not use if damaged.
  • Баалангандан ашык колдонбоңузtage between terminals or between any terminal and ground.
  • Том менен иштөөдө этият болуңузtag30V AC RMS, 42V пик же 60V DC жогору. Бул томtagшок коркунучун жаратат.
  • Иштөө алдында батарея отсеги бекем жабылганын текшериңиз.
  • Do not operate the instrument in explosive atmospheres.
  • Refer to the specifications section for maximum input/output limits.

3. Пакеттин мазмуну

Таңгактан чыгаргандан кийин, төмөндө көрсөтүлгөн нерселердин бардыгы бар экенин жана бузулбаганын текшериңиз:

Packaging accessories including the calibrator, test leads, and manual.

Figure 3.1: Packaging Accessories

  1. Таңгактоочу куту
  2. Victor 01S Temperature Signal Generator
  3. Alligator Clips
  4. Industrial Test Leads (Red and Black)
  5. Latch Key
  6. Термопара адаптер
  7. User Manual
  8. Продукциянын квалификация сертификаты

4. Продукт бүтүп калдыview

The Victor 01S features a compact design with a high-definition large LCD screen for clear readings. It includes various functions such as flashlight lighting, low battery pressure tips, automatic shut-down, and a clear zero function. The device supports both Thermocouple (TC) and Thermal Resistance (RTD) functions, along with unit conversion for temperature.

Фронт view of Victor 01S showing LCD, buttons, and input/output terminals.

Figure 4.1: Victor 01S Front View Негизги өзгөчөлүктөрү менен

4.1 Панелдик киришүү

The front panel of the Victor 01S is designed for intuitive operation. Refer to the diagram below for component identification:

Detailed diagram of the Victor 01S panel with numbered labels for each button and terminal.

Figure 4.2: Victor 01S Panel Layout

  1. LCD дисплей экран
  2. Turn temperature compensation on/off (RJ-ON button)
  3. Күйгүзүү/өчүрүү баскычы
  4. Switch output function (FUNC button)
  5. Select output range (RANGE button)
  6. Output setting button (Up/Down/Left/Right arrows)
  7. Analog output resistance low-end terminal (3W)
  8. 3W: 3-wire resistance measurement SENSE terminal
  9. 4W: 4-wire resistance measurement SENSE terminal
  10. Analog output resistance high-end terminal (4W)
  11. Select temperature unit (°C/°F button)
  12. Арткы жарык/чырак баскычы
  13. ZERO баскычы
  14. Конфигурация баскычы (CONFIG)
  15. Тtage output (+) output terminal (V)
  16. Тtage output common (-) (return) terminal (COM)

5. Орнотуу

5.1 Батареяны орнотуу

The Victor 01S is powered by three 1.5V AA batteries. To install or replace batteries:

  1. Аппараттын өчүрүлгөнүн текшериңиз.
  2. Аспаптын артындагы батарея бөлүкчөсүн табыңыз.
  3. Use a suitable tool (e.g., a coin) to unlock and open the battery compartment cover.
  4. Insert three 1.5V AA batteries, observing the correct polarity (+/-) as indicated inside the compartment.
  5. Батарея уясынын капкагын жаап, бекитиңиз.
Артка view of Victor 01S showing the battery compartment for three AA batteries.

Сүрөт 5.1: Батарея бөлүмү

Эскертүү: The instrument features low pressure tips to indicate when batteries need replacement.

5.2 Баштапкы күйгүзүү

Press the Power On/Off button (3) to turn on the device. The LCD screen (1) will illuminate, displaying the current settings.

6. Операция

The Victor 01S can output various signals. Use the FUNC button (4) to cycle through different output modes (DCV, Ohm, TC, RTD). Use the RANGE button (5) to select the appropriate measurement range within each mode. The arrow buttons (6) are used to adjust the output value.

6.1 DC Voltagд

To output a DC voltage сигнал:

  1. Press FUNC (4) until 'DCV' mode is selected.
  2. Connect the test leads to the V (+) terminal (15) and COM (-) terminal (16).
  3. Use the RANGE button (5) to select the desired voltage диапазону.
  4. Чыгаруу көлөмүн тууралаңызtage using the arrow buttons (6).

6.2 Resistance (Ohm) Output

To output a resistance signal:

  1. Press FUNC (4) until 'Ohm' mode is selected.
  2. Connect the test leads to the appropriate resistance terminals (3W, 4W, or Ω/COM) based on the required measurement type (2-wire, 3-wire, or 4-wire).
  3. Use the RANGE button (5) to select the desired resistance range.
  4. Adjust the output resistance using the arrow buttons (6).

6.3 Thermocouple (TC) Output

The Victor 01S supports 11 types of thermocouples (R, S, K, E, J, T, N, B, A, C, D).

Diagram showing 11 types of thermocouples (R, S, K, E, J, T, N, B, A, C, D) and their temperature ranges.

Figure 6.1: Supported Thermocouple Types and Ranges

  1. Press FUNC (4) until 'TC' mode is selected.
  2. Connect the thermocouple adapter to the V (+) and COM (-) terminals.
  3. Connect the thermocouple to the adapter.
  4. Use the CONFIG button (14) to select the desired thermocouple type (e.g., K, J, T).
  5. Use the RANGE button (5) to select the appropriate temperature range.
  6. Adjust the output temperature using the arrow buttons (6).
  7. Use the °C/°F button (11) to switch between Celsius and Fahrenheit units.
  8. The RJ-ON button (2) can be used to turn temperature compensation on/off.

6.4 Thermal Resistance (RTD) Output

The Victor 01S can simulate 5 types of thermal resistance (Pt100, Pt200, Pt500, Pt1000, Cu50).

Diagram showing 5 types of thermal resistance (RTD) sensors (Pt100, Pt200, Pt500, Pt1000, Cu50) and their temperature ranges.

Figure 6.2: Supported RTD Types and Ranges

  1. Press FUNC (4) until 'RTD' mode is selected.
  2. Connect the test leads to the appropriate resistance terminals (3W or 4W) based on the RTD type.
  3. Use the CONFIG button (14) to select the desired RTD type (e.g., Pt100, Cu50).
  4. Use the RANGE button (5) to select the appropriate temperature range.
  5. Adjust the output temperature using the arrow buttons (6).
  6. Use the °C/°F button (11) to switch between Celsius and Fahrenheit units.

6.5 Башка функциялар

  • ZERO Function (13): Press the ZERO button to clear the current reading or set a reference point.
  • Backlight/Flashlight (12): Press the backlight/flashlight button to illuminate the screen or use the built-in flashlight for working in low-light conditions.
  • Автоматтык түрдө өчүрүү: The instrument features an automatic shut-down function to conserve battery life.

7. Техникалык шарттар

The Victor 01S Temperature Source Signal Generator offers high precision and a wide range of measurement capabilities. All specifications are based on a calibration performed within one year, at 23°C ±5°C, 20-70% RH, with an accuracy of ± (% set value + reading).

Table of detailed specifications for DCV, Ohm, Thermocouple, and RTD output functions, including measurement range, output range, resolution, and accuracy.

Сүрөт 7.1: Толук спецификациялар таблицасы

Analog Output Function Specifications
Чыгуу функциясыӨлчөө диапазонуOutput диапазонуРезолюцияТактык
DCV100мВ-10.000мВ ~ 110.00мВ0.01мВ0.05%+0.03мВ
1V-0.1000V ~ 1.1000V0.0001V0.05%+0.3мВ
OMH400Ω0 ~ 400.0Ω0.1Ω0.05%+0.2Ω
4KΩ0 ~ 4.000КΩ0.05%+2Ω
TCR0°C ~ 1767°C1°C0°C-100°C: 1.5°C
100°C-1767°C: 1.2°C
S0°C ~ 1767°C1°C0°C-100°C: 0.9°C
100°C-1767°C: 0.9°C
K-200.0°C ~ 1372.0°C0.1°C-200°C-0°C: 1.0°C
0°C-1300°C: 0.8°C
1300°C-1372°C: 1.0°C
E-200.0°C ~ 1000.0°C0.1°C-200°C-0°C: 1.0°C
0°C-800°C: 0.8°C
800°C-1000°C: 1.0°C
J-200.0°C ~ 1200.0°C0.1°C-200°C-0°C: 1.0°C
0°C-1200°C: 0.8°C
T-250.0°C ~ 400.0°C0.1°C-250°C-0°C: 1.0°C
0°C-400°C: 0.6°C
N-200.0°C ~ 1300.0°C0.1°C-200°C-0°C: 1.0°C
0°C-1300°C: 0.8°C
B600°C ~ 1820°C1°C600°C-800°C: 1.0°C
800°C-1820°C: 0.9°C
A0°C ~ 2500°C1°C0°C-1500°C: 2.0°C
1500°C-2500°C: 2.0°C
C0°C ~ 2310°C1°C0°C-1600°C: 2.0°C
1600°C-2310°C: 2.0°C
D0°C ~ 2310°C1°C0°C-270°C: 2.0°C
270°C-2310°C: 2.0°C
RTDPt100(385)-200°C ~ 800°C0.1°C-200°C-0°C: 0.3°C
0°C-600°C: 0.2°C
600°C-800°C: 0.3°C
Pt200(385)-200°C ~ 630°C0.1°C-200°C-0°C: 0.3°C
0°C-630°C: 0.2°C
Pt500(385)-200°C ~ 630°C0.1°C-200°C-0°C: 0.3°C
0°C-630°C: 0.2°C
Pt1000(385)-200°C ~ 630°C0.1°C-200°C-0°C: 0.3°C
0°C-630°C: 0.2°C
Cu50-50°C ~ 150°C0.1°C-50°C-0°C: 0.8°C
0°C-150°C: 0.6°C

Эскертүү: Temperature coefficient: 0.1 × Basic accuracy/°C (temperature range <18°C or >28°C).

Колдонулган максималдуу томtage at output terminal: approximately 30Vpk.

Maximum applied current at output terminal: approximately 25mA.

Load characteristics: capacitive load ≥0.01uF (DCV/OHM/TC/RTD).

8. Техникалык тейлөө

8.1 Тазалоо

Аспапты тазалоо үчүн, корпусту жарнама менен сүртүңүзamp cloth and a mild detergent. Do not use abrasives or solvents. Ensure the instrument is completely dry before use.

8.2 Батареяны алмаштыруу

Replace batteries when the low battery indicator appears on the display to ensure accurate measurements and proper functionality. Refer to Section 5.1 for battery installation instructions.

8.3 Сактоо

When not in use for extended periods, remove the batteries to prevent leakage and store the instrument in a cool, dry place away from direct sunlight and extreme temperatures.

9. Кыйынчылыктарды

  • No display or erratic readings: Check battery levels and replace if necessary. Ensure test leads are securely connected.
  • Туура эмес окуулар: Verify the correct output function and range are selected. Ensure proper connection of sensors/probes. Consider recalibration if accuracy issues persist after checking connections and settings.
  • Чыгуу жок: Confirm the device is powered on and the correct output function is selected. Check for any error messages on the display.

10. Колдонуучунун кеңештери

  • The versatility of this signal generator means its 'capacity' depends entirely on the specific measurement type you are performing (e.g., resistance, thermocouple, or thermal resistance).
  • Remember that the device supports a wide array of thermocouple types (R, S, K, E, J, T, N, B, A, C, D) and RTD types (Pt100, Pt200, Pt500, Pt1000, Cu50). Always select the correct type in the settings for accurate simulation.
  • For precise measurements, especially with resistance, utilize the 4-wire connection method when possible to minimize lead resistance errors.

11. Manufacturing Process

The Victor 01S signal generator is manufactured using advanced techniques, including automated circuit board assembly to ensure high quality and reliability.

Video 11.1: Manufacturing process of circuit boards for the Victor 01S.

12. Колдоо

For technical assistance, troubleshooting beyond this manual, or inquiries regarding your Victor 01S Temperature Source Signal Generator, please contact your local distributor or the manufacturer's customer support. Refer to the product packaging or qualification certificate for contact details.

Тектеш документтер - 01S

Preview VICTOR 4105A Санариптик жерге туташтыруу каршылык өлчөгүч инструкция
VICTOR 4105A Санариптик Жер Сыноочу үчүн комплекстүү нускамаview, коопсуздук чаралары, иш принциби, техникалык мүнөздөмөлөр, өлчөө ыкмалары, батареяны орнотуу жана көйгөйлөрдү чечүү.
Preview VICTOR электрондук сыноо жана өлчөө приборлор каталогу | Санариптик мультиметрлер, осциллографтар жана башкалар
VICTOR электрондук сыноо жана өлчөө приборлорунун комплекстүү спектрин изилдеңиз, анын ичинде санариптик мультиметрлер, cl.amp метрлер, осциллографтар, функциялык генераторлор жана башкалар. Кесиптик колдонуу үчүн жогорку сапаттагы, ишенимдүү жабдуулар.
Preview Victor 1226 калькуляторун тез баштоо боюнча колдонмо
Victor 1226 калькуляторун орнотуу, кагаз жүктөө, убакытты жана датаны коюу, мүчүлүштүктөрдү оңдоо жана жеткиликтүү аксессуарларды камтыган кыскача колдонмо.
Preview Victor Personal Massager Колдонуучу колдонмосу жана Joyhub колдонмосу
Victor жеке массажери үчүн комплекстүү колдонуучу колдонмосу, өнүм өзгөчөлүктөрү, Joyhub менен колдонмо туташуу, колдонуу нускамалары, кам көрүү, эскертүүлөр жана кепилдик маалыматы. Бир нече тилди колдойт.
Preview Victor 1226 Thermal Printing Calculator Instruction Manual
This manual provides instructions for operating the Victor 1226 Thermal Printing Calculator, covering its features, functions, and specifications. It details preparation, key identification, date and time functions, battery replacement, and various calculation examples.
Preview Victor Digital Anemometer Instruction Manual
Instruction manual for the Victor Digital Anemometer, detailing its functions, LCD display, operation, and specifications. Covers air velocity and temperature measurements, data hold, unit selection, and power management.