1. Киришүү
The Victor 01S is a high-precision temperature source signal generator designed for calibrating and testing various temperature sensors and electrical signals. It features a basic accuracy of 0.05% and a 5-digit output display, making it suitable for professional and industrial applications. This manual provides detailed instructions for the safe and effective use of your device.

2. Коопсуздук маалыматы
Please read and understand all safety warnings and operating instructions before using this device. Failure to do so may result in injury or damage to the instrument.
- Түзмөктү ар дайым көрсөтүлгөн көлөмдө колдонуңузtage жана учурдагы чектөөлөр.
- Do not operate the device in wet environments or near explosive gases.
- Ensure proper connection of test leads to avoid electrical shock or damage.
- Replace batteries promptly when the low pressure tips indicator appears.
- Do not attempt to repair or modify the device yourself. Refer servicing to qualified personnel.
3. Продукт бүтүп калдыview
3.1. Панелге киришүү

- LCD дисплей экран
- Turn temperature compensation on/off (RJ-ON button)
- Power on/off (Green button)
- Switch output function (FUNC button)
- Select output range (RANGE button)
- Output setting button (CONFIG button)
- Analog output resistance low-end terminal
- 3W: 3-wire resistance measurement terminal
- 4W: 4-wire resistance measurement SENSE terminal
- Analog output resistance high-end terminal
- Select temperature unit (°C/°F button)
- Backlight/flashlight button (Yellow button)
- ZERO баскычы
- Configuration button (Arrow keys)
- V: Тtage output (+) output terminal
- COM: Тtage output common (-) (return) terminal
3.2. Негизги өзгөчөлүктөрү
- HD large screen display for clear readings.
- Flashlight lighting for use in low-light conditions.
- Low pressure tips for battery status indication.
- Automatic shut-down to conserve battery life.
- Clear function (ZERO button).
- Thermocouple (TC) function.
- Thermal resistance (RTD) function.
- Unit conversion (°C/°F).
- High-precision measurement with 0.05% accuracy and 5-digit output.

3.3. Пакеттин мазмуну

- Таңгактоочу куту
- Temperature Signal Generator (Victor 01S)
- Аллигатор клиптери
- Өнөр жай сыноолору
- Latch key
- Thermocouple adapter
- User Manual
- Продукциянын квалификациясынын сертификаты
4. Орнотуу
4.1. Батареяны орнотуу
The Victor 01S requires three (3) 1.5V AA batteries for operation. To install or replace batteries:
- Аппараттын артындагы батарея бөлүкчөсүн табыңыз.
- Батареянын капкагын ачыңыз.
- Insert three 1.5V AA batteries, ensuring correct polarity (+/-).
- Батареянын капкагын бекем жабыңыз.

4.2. Сыноо линияларын туташтыруу
Connect the industrial test leads or alligator clips to the appropriate terminals on the device based on the desired measurement or output function (e.g., V, COM, 3W, 4W, Ω). Ensure connections are firm and secure.
5. Иштөө нускамалары
5.1. Негизги операция
- Күйгүзүү/өчүрүү: Press the green power button (3) to turn the device on or off.
- Функция тандоо: Press the 'FUNC' button (4) to cycle through different output functions (e.g., DCV, OMH, TC, RTD).
- Аралыкты тандоо: Use the 'RANGE' button (5) to select the appropriate measurement or output range for the selected function.
- Чыгуу жөндөөлөрү: Use the 'CONFIG' button (6) and arrow keys (14) to adjust the output value.
- Нөл функциясы: Press the 'ZERO' button (13) to clear or zero the current reading/output.
- Бирдикти айландыруу: Press the '°C/°F' button (11) to switch between Celsius and Fahrenheit temperature units.
- Арткы жарык/чырак: Press the yellow button (12) to activate the display backlight or the flashlight.
5.2. Thermocouple (TC) Simulation
The Victor 01S can simulate 11 types of thermocouple signals. Connect the thermocouple adapter and test leads to the appropriate terminals.

- R-type: 0°C ~ 1767°C
- S түрү: 0°C ~ 1767°C
- K-type: -200°C ~ 1372°C
- E-type: -200°C ~ 1000°C
- J-type: -200°C ~ 1200°C
- T-type: -250°C ~ 400°C
- N-түрү: -200°C ~ 1300°C
- B-type: 600°C ~ 1820°C
- A-type: 0°C ~ 2500°C
- C-type: 0°C ~ 2310°C
- D-type: 0°C ~ 2310°C
5.3. Thermal Resistance (RTD) Simulation
The device can output 5 types of RTD signals. Connect test leads to the 3W or 4W terminals as required.

- Pt100: -200°C ~ 800°C
- Cu50: -50°C ~ 150°C
- Pt200: -200°C ~ 630°C
- Pt1000: -200°C ~ 630°C
- Pt500: -200°C ~ 630°C
6. Техникалык шарттар
The following table details the analog output functions and their respective ranges, resolutions, and accuracies. Specifications are valid for one year after calibration, at 23°C ±5°C, 20-70% RH, accuracy = ± (% set value + reading).

| Output Function | Өлчөө диапазону | Чыгуу диапазону | Резолюция | Тактык |
|---|---|---|---|---|
| DCV | -10.000мВ ~ 110.00мВ | 0 ~ 1.1000V | 0.01мВ | 0.05% + 0.03мВ |
| DCV | 1V | -0.1000V ~ 1.1000V | 0.0001V | 0.05% + 0.2мВ |
| OMH | 400Ω | 0 ~ 400.0Ω | 0.1Ω | 0.05% + 2Ω |
| OMH | 4KΩ | 0 ~ 4.000КΩ | 1Ω | 0.05% + 20Ω |
| TC (R) | 0°C ~ 1767°C | 0°C ~ 1767°C | 1°C | 0°C ~ 100°C: 1.5°C 100°C ~ 1767°C: 1.2°C |
| TC (S) | 0°C ~ 1767°C | 0°C ~ 1767°C | 1°C | 0°C ~ 100°C: 1.5°C 100°C ~ 1767°C: 1.2°C |
| TC (K) | -200°C ~ 1372°C | -200°C ~ 1372°C | 0.1°C | -200°C ~ 0°C: 0.9°C 0°C ~ 1000°C: 0.6°C 1000°C ~ 1372°C: 0.7°C |
| TC (E) | -200°C ~ 1000°C | -200°C ~ 1000°C | 0.1°C | -200°C ~ 0°C: 0.8°C 0°C ~ 800°C: 0.5°C 800°C ~ 1000°C: 0.6°C |
| TC (J) | -200°C ~ 1200°C | -200°C ~ 1200°C | 0.1°C | -200°C ~ 0°C: 0.8°C 0°C ~ 1000°C: 0.5°C 1000°C ~ 1200°C: 0.6°C |
| TC (T) | -250.0°C ~ 400.0°C | -250.0°C ~ 400.0°C | 0.1°C | -250.0°C ~ -100°C: 0.6°C -100°C ~ 400°C: 0.5°C |
| TC (N) | -200°C ~ 1300°C | -200°C ~ 1300°C | 0.1°C | -200°C ~ 0°C: 0.8°C 0°C ~ 1000°C: 0.6°C 1000°C ~ 1300°C: 0.7°C |
| TC (B) | 600°C ~ 1820°C | 600°C ~ 1820°C | 1°C | 600°C ~ 1500°C: 2.0°C 1500°C ~ 1820°C: 2.4°C |
| TC (A) | 0°C ~ 2500°C | 0°C ~ 2500°C | 1°C | 0°C ~ 1600°C: 2.0°C 1600°C ~ 2500°C: 3.0°C |
| TC (C) | 0°C ~ 2310°C | 0°C ~ 2310°C | 1°C | 0°C ~ 270°C: 2.5°C 270°C ~ 2310°C: 3.0°C |
| TC (D) | 0°C ~ 2310°C | 0°C ~ 2310°C | 1°C | 0°C ~ 270°C: 2.5°C 270°C ~ 2310°C: 3.0°C |
| RTD (Pt100(385)) | -200°C ~ 800°C | -200°C ~ 800°C | 0.1°C | -200°C ~ 0°C: 0.3°C 0°C ~ 600°C: 0.2°C 600°C ~ 800°C: 0.3°C |
| RTD (Pt200(385)) | -200°C ~ 630°C | -200°C ~ 630°C | 0.1°C | -200°C ~ 0°C: 0.3°C 0°C ~ 630°C: 0.2°C |
| RTD (Pt500(385)) | -200°C ~ 630°C | -200°C ~ 630°C | 0.1°C | -200°C ~ 0°C: 0.3°C 0°C ~ 630°C: 0.2°C |
| RTD (Pt1000(385)) | -200°C ~ 630°C | -200°C ~ 630°C | 0.1°C | -200°C ~ 0°C: 0.3°C 0°C ~ 630°C: 0.2°C |
| RTD (Cu50) | -50°C ~ 150°C | -50°C ~ 150°C | 0.1°C | -50°C ~ 150°C: 0.6°C |
6.1. Жалпы спецификациялар
- Өлчөөнүн тактыгы: 0.05%
- Батарея түрү: 3 x 1.5V AA батарейкалар (киргизилген эмес)
- Maximum Applied Voltage Чыгуу терминалында: Болжол менен 30Vpk
- Maximum Applied Current at Output Terminal: Болжол менен 25мА
- Load Characteristics: Capacitive load ≥0.01uF (DCV/OHM/TC/RTD)
- Температура коэффициенти: 0.1 × Basic accuracy/°C (Temperature range <18°C or >28°C)
- Өлчөмдөрү (L x W x Y): 24 см x 19 см x 7 см (болжолдуу пакеттин өлчөмдөрү)
- Салмагы: 0.799 кг (болжолдуу пакеттин салмагы)
7. Техникалык тейлөө
- Тазалоо: жумшак колдонуу, damp Түзмөктү тазалоо үчүн чүпүрөктү колдонуңуз. Абразивдүү тазалоочу каражаттарды же эриткичтерди колдонбоңуз.
- Батареяны алмаштыруу: Replace batteries when the low battery indicator appears to ensure accurate readings and prevent data loss.
- Сактагыч: Түзмөктү салкын, кургак жерде, күндүн түз нурунан жана өтө жогорку температурадан алыс сактаңыз. Узак убакытка сактасаңыз, батареяларды алып салыңыз.
- Калибрлөө: For optimal accuracy, periodic calibration by a qualified service center is recommended, especially after one year of use.
8. Кыйынчылыктарды
- Түзмөк күйбөйт: Батареянын орнотулушун текшерип, батарейкалары түгөнүп калбагандыгын текшериңиз.
- Туура эмес окуулар: Verify correct function and range selection. Ensure test leads are properly connected and not damaged. Consider recalibration if issues persist.
- Дисплей күңүрт: Activate the backlight using the yellow button. If still dim, replace batteries.
- Чыгуу сигналы жок: Check output settings and connections. Ensure the device is in the correct output mode.
9. Колдонуучунун кеңештери
Ал эми re колдонуучусунун конкреттүү кеңештериviews or Q&A are not available, here are some general recommendations for using the Victor 01S:
- Always double-check your connections before applying power or generating a signal to prevent damage to the device or the circuit under test.
- Familiarize yourself with the different thermocouple and RTD types and their respective temperature ranges to ensure you select the correct setting for your application.
- Utilize the automatic shut-down feature to prolong battery life, especially during prolonged periods of inactivity.
- Keep the device clean and free from dust and moisture to maintain its performance and longevity.
10. Manufacturing Process Video
Watch this video to see the manufacturing and quality control processes for Victor products, ensuring high standards and reliability.
11. Кепилдик жана колдоо
For warranty information, technical support, or service inquiries, please contact your authorized Victor dealer or the manufacturer directly. Keep your purchase receipt and product qualification certificate for reference.