Oakton WD-35805-04 Laboratory pH Electrode Instruction Manual

Model: WD-35805-04

Introduction

This manual provides comprehensive instructions for the proper setup, operation, and maintenance of your Oakton WD-35805-04 Laboratory pH Electrode. Adhering to these guidelines will ensure optimal performance and longevity of the device.

Продукт бүттүview

The Oakton WD-35805-04 is a double-junction, glass body, refillable pH electrode designed for laboratory applications. Its robust construction and double-junction design make it suitable for a wide range of chemical measurements, offering enhanced durability and resistance to aggressive chemicals. This electrode features an Ag/AgCl reference cell for stable and accurate readings.

Oakton WD-35805-04 Laboratory pH Electrode

Image 1: Oakton WD-35805-04 Laboratory pH Electrode. This image shows the full electrode, featuring a black top housing with an "OPEN" indicator, a clear glass body revealing internal components, and a blue spherical tip. A black cable extends from the top.

Негизги өзгөчөлүктөрү:

Жайгашуу

  1. Пакеттен чыгаруу: Carefully remove the electrode from its packaging. Inspect for any visible damage.
  2. Electrolyte Filling (if necessary): This electrode is refillable. If the electrolyte level is low, open the fill hole (usually indicated by an "OPEN" arrow or cap) and carefully add the appropriate reference electrolyte solution (e.g., 3M KCl) until the level is just below the fill hole. Close the fill hole securely.
  3. Чайкоо: Rinse the electrode tip with distilled or deionized water to remove any storage solution. Do not wipe the glass bulb, as this can create static charges.
  4. Кондициялоо: For optimal performance, soak the electrode in a pH 7 buffer solution for at least 30 minutes (or longer, up to several hours) before first use or after prolonged dry storage. This rehydrates the glass membrane.
  5. Connecting to Meter: Connect the electrode's BNC connector (or other specified connector type) to the corresponding input on your pH meter. Ensure a secure connection.
  6. Калибрлөө: Calibrate your pH meter with the electrode using at least two, preferably three, fresh pH buffer solutions (e.g., pH 4.01, 7.00, and 10.00). Follow your pH meter's specific calibration procedure.

Иштөө нускамалары

  1. Даярдоо: Ensure the electrode is properly conditioned and calibrated.
  2. Sample өлчөө:
    • с ортосунда дистилденген же деионизацияланган суу менен электродду чайкаңызamples.
    • Электроддун учу s чөмүлөample solution, ensuring the glass bulb and junction are fully submerged.
    • Stir the solution gently to ensure homogeneity and faster response.
    • Allow the reading to stabilize before recording the pH value. Stability is typically indicated by a stable reading on the pH meter for a few seconds.
  3. Температураны компенсациялоо: If your pH meter has automatic temperature compensation (ATC), ensure the temperature probe is also immersed in the sample. If not, manually input the sample temperature into the meter.
  4. Post-Measurement: After each measurement, rinse the electrode thoroughly with distilled or deionized water.

Техникалык тейлөө

Проблемаларды чечүү

КөйгөйМүмкүн себепЧечим
Slow Response / Unstable Readings
  • Dehydrated glass membrane
  • Тосулган түйүн
  • Төмөн электролит деңгээли
  • Contaminated buffer solutions
  • Температура туруктуу эмес
  • Recondition electrode in pH 7 buffer.
  • Clean electrode as per Maintenance section.
  • Refill electrolyte.
  • Жаңы эритмелерди колдонуңуз.
  • уруксат берүү sample/electrode to reach thermal equilibrium.
Туура эмес окуулар
  • Туура эмес калибрлөө
  • Кир электрод
  • Expired or contaminated buffers
  • Electrode nearing end of life
  • Recalibrate with fresh buffers.
  • Clean electrode.
  • Replace buffers.
  • Consider replacing electrode if issues persist after cleaning/reconditioning.
No Reading / "Error" on Meter
  • Бошоң байланыш
  • Сынган электрод
  • Эсептегичтин бузулушу
  • Check electrode connection to meter.
  • Inspect electrode for visible damage (cracks in glass).
  • Test meter with a known good electrode or consult meter manual.

Техникалык шарттар

АтрибутДеталь
Модель номериWD-35805-04
Электрод түрүLaboratory pH Electrode, Double-Junction
Дене материалыАйнек
Шилтеме түрүRefillable Ag/AgCl
Пакеттин өлчөмдөрү10.9 x 4.1 x 2.2 дюйм
Элемент Салмагы6.38 унция
ӨндүрүүчүOAKTON Instruments
Дата биринчи жеткиликтүү2-июнь, 2016-жыл
ASINB01GIRDK64
Best Sellers RankӨнөр жай жана илимий чөйрөдөгү мыкты 100дү караңыз

Кепилдик жана колдоо

For specific warranty information regarding your Oakton WD-35805-04 Laboratory pH Electrode, please refer to the documentation included with your purchase or visit the official Oakton Instruments website. For technical support, troubleshooting assistance beyond this manual, or to inquire about replacement parts and accessories, please contact Oakton Instruments customer service directly.

Oakton Instruments Contact Information: (Сураныч, продуктунун таңгагын же расмий website for current contact details.)

Тектеш документтер - WD-35805-04

Preview Ойготкуч менен OAKTON 410 Таймер: Иштөө нускамалары жана спецификациялар
Ойготкучтуу OAKTON 410 Таймери (WD-35002-00 модели) үчүн комплекстүү иштетүү нускамалары жана спецификациялары, анын өзгөчөлүктөрү, баскыч функциялары, убакыт режимдери жана ойготкуч орнотуулары.
Preview OAKTON TEMP TC Acorn Series Thermocouple Thermometer Operating Instructions
Operating instructions and specifications for the OAKTON TEMP TC Acorn Series Thermocouple Thermometer, including details on measurement, calibration, troubleshooting, and accessories.
Preview OAKTON ION Acorn Series Meters: Operating Instructions and Guide
Comprehensive operating manual for OAKTON ION 5 and ION 6 Acorn Series pH/Ion Meters. Covers setup, calibration, measurement, maintenance, and troubleshooting for accurate water quality analysis.
Preview OAKTON Single Event 220 Секундомер Сигналдын иштөө инструкциясы менен
Ойготкучтуу OAKTON Single-Event 220 секундомер (WD-35002-10 модели) үчүн толук иштетүү нускамалары, секундомер функцияларын, убакыт/календар орнотууларын, ойготкучтун өзгөчөлүктөрүн, спецификацияларын жана кепилдик маалыматын камтыйт.
Preview Октон DO250/DO260 эриген кычкылтек өлчөгүч инструкциясы
Oakton DO250 жана DO260 эриген кычкылтек өлчөгүчтөрү үчүн комплекстүү нускамаview, негизги операциялар, калибрлөө, маалыматтарды башкаруу, орнотуу, тейлөө, бузулууларды аныктоо жана коопсуздук чаралары.
Preview OAKTON 300 Series Handheld Water Quality Meters Instruction Manual
This instruction manual provides comprehensive guidance for operating the OAKTON 300 series handheld water quality meters. It covers product overview, setup, calibration, measurement, data management, maintenance, and troubleshooting for accurate water quality analysis.