The INSTRUKART AI-DP1-LCD is a digital differential pressure indicator designed for precise measurement in critical environments such as laminar airflow cabinets, clean rooms, biosafety cabinets, and Air Handling Units (AHU). This device serves as a modern replacement for traditional analog Magnehelic gauges, offering enhanced readability and functionality. It features a clear Graphic Backlit 128 x 64 LCD, an integrated high-speed differential pressure sensor, and a built-in buzzer alarm for process violation alerts with an acknowledgment key.
This manual provides essential information for the safe and efficient operation, installation, and maintenance of your AI-DP1-LCD Differential Pressure Indicator. Please read this manual thoroughly before using the device.
2. Коопсуздук маалыматы
Always adhere to local safety regulations and guidelines when installing and operating the device. Ensure that the power supply matches the specified requirements (24V DC). Do not expose the device to corrosive gases or liquids. Only qualified personnel should perform installation and maintenance.
3. Пакеттин мазмуну
Таңгактан чыгаргандан кийин, төмөндө көрсөтүлгөн нерселердин бардыгы бар экенин жана бузулбаганын текшериңиз:
- 1 Unit of INSTRUKART AI-DP1-LCD Differential Pressure Indicator
- Кубаттуу адаптер
- Көрсөтмө (бул документ)
- Factory Calibration Certificate with NABL Standards
4. Продукт бүтүп калдыview
The AI-DP1-LCD is designed for accurate differential pressure monitoring. Its robust construction and advanced features make it suitable for demanding industrial and scientific applications.
4.1 Негизги өзгөчөлүктөрү
- Digital readout with Graphic Backlit 128 x 64 LCD.
- Integrated high-speed differential pressure sensor.
- Built-in high and low buzzer alarm for process set parameter violation with acknowledgment key.
- Unit conversion capability (mm.w.c. & Pascals) - Note: Unit selection is factory fixed for this model (-180 to 180 Pascals).
- Offset provision to meet audit requirements.
- 30 mm depth for cleanroom modular wall fitment.
- Isolated RS485 Modbus communication for BMS / SCADA / PLC integration.
4.2 Продукт Views

1-сүрөт: Бурчтуу view of the INSTRUKART AI-DP1-LCD Differential Pressure Indicator.

Сүрөт 2: Алдыңкы view of the AI-DP1-LCD, showing the display and 'ACK' button.

Figure 3: Comparison of the side and front views, highlighting the slim profile and rear connections.
5. Орнотуу жана орнотуу
The AI-DP1-LCD is designed for modular wall fitment in cleanroom environments.
5.1 Монтаждоо
The instrument enclosure has dimensions of 100mm (W) × 90mm (H) × 30mm (D), allowing for flush mounting into cleanroom modular walls. Ensure adequate space for rear connections.
5.2 Басым байланыштары
Connect the pressure tubing to the two 4mm stainless steel hose nipples located at the back of the unit. Ensure connections are secure to prevent leaks. The device is designed for use with air or non-corrosive gases.
5.3 Кубат туташтыруу
Connect the provided power adaptor to the device's power input. The typical power requirement is 24V DC.
5.4 Communication (Optional)
For integration with Building Management Systems (BMS), SCADA, or PLC, utilize the isolated RS485 communication port with Modbus protocols. Refer to the Modbus communication protocol documentation for detailed setup instructions (not included in this manual).
6. Иштөө нускамалары
6.1 Кубат күйгүзүлдү
Once powered on, the Graphic Backlit LCD will display the current differential pressure reading. The display also shows the date and time.
6.2 Reading Display
The LCD shows the differential pressure value in Pascals. This model has a factory-fixed unit selection for Pascals.
6.3 Ойготкуч функциясы
The device features a built-in high and low buzzer alarm. If the measured differential pressure exceeds or falls below the pre-set parameters, the buzzer will activate to indicate a process violation.
6.4 Alarm Acknowledgment
To acknowledge and silence an active alarm, press the 'ACK' button located on the front panel below the display. Acknowledging the alarm does not resolve the underlying issue; it only silences the buzzer.
7. Техникалык тейлөө
7.1 Тазалоо
Wipe the front stainless steel plate and LCD screen with a soft, damp cloth. Do not use abrasive cleaners or solvents, as these may damage the surface or display. Ensure no liquid enters the device.
7.2 Калибрлөө
The device comes with a Factory Calibration Certificate with NABL Standards. Regular re-calibration may be required depending on application and regulatory requirements. Contact INSTRUKART for calibration services.
8. Кыйынчылыктарды
| Көйгөй | Мүмкүн себеп | Чечим |
|---|---|---|
| No display/Device not powering on | Электр энергиясы жок же туура эмес томtage. | Check power adaptor connection and ensure 24V DC supply. |
| Туура эмес басым окуу | Leaky pressure connections or blocked hose nipples. | Verify all hose connections are secure. Check for any blockages in the tubing or nipples. |
| Сигнализация тынымсыз иштеп турат | Differential pressure outside set parameters. | Investigate the environmental conditions causing the pressure deviation. Acknowledge the alarm by pressing 'ACK'. |
| Communication issues (RS485) | Incorrect wiring or Modbus configuration. | Verify RS485 wiring and Modbus parameters. Consult Modbus documentation or INSTRUKART support. |
9. Техникалык шарттар
| Параметр | Деталь |
|---|---|
| Өлчөө диапазону | -180 to 180 Pascals |
| Тактык | Толук масштабдын ±0.5% |
| Дисплей | Graphic Backlit 128 x 64 LCD |
| Maximum Over Pressure | 4 Times the rated pressure |
| Сенсор | Integrated High Speed Differential Pressure Sensor |
| Электр камсыздоо | 24V DC Typical |
| Ойготкучтар | Built-in High & Low Buzzer Alarm with Acknowledgment Key |
| Курулуш | M.S. Powder Coated Body (Back) with Stainless Steel Front Flush Plate |
| Байланыш | Isolated RS485 communication, Modbus Protocols |
| Басым туташуу | 2 x 4mm Stainless Steel Hose Nipples at the back |
| Орто | Air or Non-Corrosive Gases |
| Өлчөмдөрү (W×H×D) | 100 мм × 90 мм × 30 мм |
| Бирдикти тандоо | Factory Fixed for Pascals |
| Элемент Салмагы | 1.76 фунт (болжол менен 0.8 кг) |
10. Тиркемелер
The INSTRUKART AI-DP1-LCD is suitable for a wide range of critical environments requiring precise differential pressure monitoring:
- Infection Rooms
- Hospital Operation Theatre (OT) rooms
- Isolation Rooms (especially where negative pressure is required)
- Laminar Air Flow Cabinets
- Pharmaceutical Clean Rooms
- HVAC системалары
- Bio Safety Cabinets
- Air Handling Units (AHUs)

Figure 4: The AI-DP1-LCD installed in an isolation room, demonstrating its application in critical healthcare environments.

Figure 5: Visual representation of diverse applications for the differential pressure indicator, including cleanrooms, HVAC, and pharmaceutical settings.
11. Кепилдик жана колдоо
For warranty information, technical support, or service inquiries, please contact INSTRUKART directly. Refer to the product packaging or INSTRUKART's official webакыркы байланыш маалыматы үчүн сайт.